Remote Commands
R&S
®
FPL1000
725
User Manual 1178.3370.02 ─ 11
<li>
Return values:
<Condition>
0
The marker has passed the limit check.
1
The marker is inside the margins of a limit line.
2
The marker has failed the limit check.
Example:
CALC:MARK1:FUNC:FME:LIM2:LCON?
Queries the condition of marker 1 in relation to limit line 2.
Usage:
Query only
CALCulate<n>:DELTamarker<m>:FUNCtion:FMEasurement:LIMit<li>:LDELta?
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:LIMit<li>:LDELta?
<Amplitude>
This command queries the vertical distance from the marker position to the limit line.
The unit is dB.
If the marker has been assigned to a different trace than the limit line, or if no limit ine
is defined for the marker position, the command returns -200.
Suffix:
<n>
.
<m>
<li>
1..n
Return values:
<Amplitude>
Vertical distance to the limit line in dB.
Example:
CALC:MARK3:FUNC:FME:LIM2:LDEL?
Queries the distance of marker 3 to the second limit line.
Usage:
Query only
10.6.14.7
Evaluating the Results
The commands required to control the demodulation of signals at the marker position
are described in
Chapter 10.8.8.14, "Marker Demodulation"
10.6.14.8
Programming Example: EMI Measurement
This example demonstrates how to detect electromagnetic interferences (EMI) in a
remote environment.
//----------- Preparing the measurement ------------
//Reset the instrument
*RST
Configuring and Performing Measurements