The Spectrum Application (RF Measurements)
R&S
®
FPL1000
282
User Manual 1178.3370.02 ─ 11
Remote command:
Meas Time/Average
Selects the measurement time or switches to manual averaging mode. In general,
results are more precise with longer measurement times. The following settings are
recommended for different signal types to obtain stable and precise results:
"Short"
Stationary signals with high power (> -40dBm), because they require
only a short measurement time and short measurement time provides
the highest repetition rates.
"Normal"
Signals with lower power or modulated signals
"Long"
Signals at the lower end of the measurement range (<-50 dBm) or
Signals with lower power to minimize the influence of noise
"Manual"
Manual averaging mode. The average count is set with the
Remote command:
[SENSe:]PMETer<p>:MTIMe:AVERage[:STATe]
Setting the Reference Level from the Measurement Meas -> Ref
Sets the currently measured power as a reference value for the relative display. The
reference value can also be set manually via the
Remote command:
CALCulate<n>:PMETer<p>:RELative[:MAGNitude]:AUTO ONCE
Reference Value
Defines the reference value in dBm used for relative power meter measurements.
Remote command:
CALCulate<n>:PMETer<p>:RELative[:MAGNitude]
Use Ref Level Offset
If activated, takes the reference level offset defined for the analyzer into account for the
measured power (see
"Shifting the Display (Offset)"
If deactivated, takes the
into account.
Remote command:
[SENSe:]PMETer<p>:ROFFset[:STATe]
Sensor Level Offset
Takes the specified offset into account for the measured power. Only available if
is disabled.
Remote command:
Receiving Data Input and Providing Data Output