![Ronde & Schwarz 1304.0004K03 User Manual Download Page 678](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987678.webp)
Remote Commands
R&S
®
FPL1000
678
User Manual 1178.3370.02 ─ 11
................................................................................... 679
[SENSe:]LIST:RANGe<ri>:DETector
.................................................................................679
[SENSe:]LIST:RANGe<ri>[:FREQuency]:STARt
................................................................ 680
[SENSe:]LIST:RANGe<ri>[:FREQuency]:STOP
.................................................................680
[SENSe:]LIST:RANGe<ri>:FILTer:TYPE
............................................................................681
[SENSe:]LIST:RANGe<ri>:INPut:ATTenuation
................................................................... 681
[SENSe:]LIST:RANGe<ri>:INPut:ATTenuation:AUTO
......................................................... 682
[SENSe:]LIST:RANGe<ri>:INPut:GAIN:STATe
...................................................................682
[SENSe:]LIST:RANGe<ri>:INPut:GAIN[:VALue]
.................................................................682
[SENSe:]LIST:RANGe<ri>:LIMit:STARt
.............................................................................682
[SENSe:]LIST:RANGe<ri>:LIMit:STATe
.............................................................................683
[SENSe:]LIST:RANGe<ri>:LIMit:STOP
............................................................................. 683
[SENSe:]LIST:RANGe<ri>:POINts[:VALue]
....................................................................... 683
................................................................................... 684
[SENSe:]LIST:RANGe<ri>:SWEep:TIME
.......................................................................... 684
[SENSe:]LIST:RANGe<ri>:SWEep:TIME:AUTO
................................................................ 684
[SENSe:]LIST:RANGe<ri>:TRANsducer
........................................................................... 685
[SENSe:]LIST:RANGe<ri>:BANDwidth:RESolution
<RBW>
This command defines the resolution bandwidth for a spurious emission measurement
range.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<RBW>
Resolution bandwidth.
Refer to the data sheet for available resolution bandwidths.
Default unit: Hz
Example:
LIST:RANG2:BAND:RES 3KHZ
Manual operation:
See
[SENSe:]LIST:RANGe<ri>:BANDwidth:VIDeo
<VBW>
This command defines the video bandwidth for a spurious emission measurement
range.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<VBW>
Video bandwidth.
Refer to the data sheet for available video bandwidths.
Default unit: Hz
Example:
LIST:RANG2:BAND:VID 3KHZ
Manual operation:
See
Configuring and Performing Measurements