Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
LIST:RANG2:SWE:TIME:AUTO ON
Manual operation:
See
[SENSe:]LIST:RANGe<ri>:TRANsducer
<Transducer>
This command selects a transducer factor for a spurious emission measurement
range.
Note the following prerequisites for the selected transducer:
●
The transducer must cover at least the span of the range.
●
The x-axis has to be linear.
●
The unit has to be dB.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<Transducer>
String containing the transducer file name. Do not include a file
extension or the file path.
The file must be located in the
C:\Users\Public\Documents\Rohde-Schwarz\Analyzer\
trd
directory.
Example:
LIST:RANG2:TRAN 'MYTRANS'
Manual operation:
See
10.6.8.3
Configuring the List Evaluation
The following commands configure the list evaluation.
Useful commands for spurious emission measurements described elsewhere
●
Remote commands exclusive to spurious emission measurements
....................................................................................... 686
..................................................................................686
CALCulate<n>:ESPectrum:PSEarch:DETails
.................................................................... 686
CALCulate<n>:ESPectrum:PEAKsearch:DETails
...............................................................686
.................................................................................... 686
CALCulate<n>:PEAKsearch:MARGin
...............................................................................686
Configuring and Performing Measurements