Remote Commands
R&S
®
FPL1000
808
User Manual 1178.3370.02 ─ 11
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>[:STATe]
<State>
This command turns a trace on and off.
The measurement continues in the background.
Suffix:
<n>
.
<w>
subwindow
Not supported by all applications
<t>
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
DISP:TRAC3 ON
Manual operation:
See
"Trace 1/Trace 2/Trace 3/Trace 4/Trace 5/Trace 6"
on page 399
See
"Trace 1/ Trace 2/ Trace 3/ Trace 4 (Softkeys)"
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:SMOothing:APERture
<Aperture>
This command defines the degree (aperture) of the trace smoothing, if
WINDow<n>][:SUBWindow<w>]:TRACe<t>:SMOothing[:STATe]
TRUE
.
Suffix:
<n>
.
<w>
subwindow
<t>
Parameters:
<Aperture>
Range:
1 to 50
*RST:
2
Default unit: PCT
Example:
DISP3:TRAC2:SMO:APER 5
Defines an aperture of 5% for trace 2 in window 3
Manual operation:
See
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:SMOothing[:STATe]
<State>
This command turns trace smoothing for a particular trace on and off.
Setting Basic Spectrum RF Measurement Parameters