Table 6-15: Attributes and child nodes of the Range element (normal ranges)
Child node
Attribute
Value
Parameter description
Mand.
Index
0…19
Indices are continuous and
have to start with 0
Yes
Name
<string>
Name of the range
Only if Referen-
ceChannel con-
tains a name and
the range is the
reference range
Short-
Name
<string>
Short name of the range
No
ChannelType
TX | Adjacent
Yes
WeightingFilter
Only if Referen-
cePower method
is TX Channel
Power and the
range is the ref-
erence range
Type
RRC | CFilter
Type of the weighting filter
Yes
Roll Off Factor
0…1
Excess bandwidth of the fil-
ter
Only if the filter
type is RRC
Bandwidth
<bandwidth in Hz>
Filter bandwidth
Only if the filter
type is RRC
FrequencyRange
Yes
Start
<frequency in Hz>
Start value of the range
Yes
Stop
<frequency in Hz>
Stop value of the range
Yes
Limit
dBm/Hz | dBm |
dBc | dBr | dB
A Range must contain
exactly two limit nodes; one
of the limit nodes has to
have a relative unit (e.g.
dBc), the other one must
have an absolute unit (e.g.
dBm)
Yes
Start
Value
<numeric_value>
Power limit at start fre-
quency
Yes
Unit
dBm/Hz | dBm |
dBc | dBr | dB
Sets the unit of the start
value
Stop
Value
<numeric_value>
Power limit at stop fre-
quency
Unit
dBm/Hz | dBm |
dBc | dBr | dB
Sets the unit of the stop
value
LimitFailMode
Absolute | Relative
| Absolute and Rel-
ative | Absolute or
Relative
If used, it has to be identical
to DefaultLimitFailMode
No
RBW
Bandwidth <bandwidth in Hz>
Yes
Measurements and Results