![Ronde & Schwarz 1304.0004K03 User Manual Download Page 393](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987393.webp)
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
393
User Manual 1178.3370.02 ─ 11
Auto detector
If the R&S
FPL1000 is set to define the appropriate detector automatically, the detector
is set depending on the selected trace mode:
Trace mode
Detector
Clear Write
Auto Peak
Max Hold
Positive Peak
Min Hold
Negative Peak
Average
Sample Peak
View
–
Blank
–
X-Value of the Sweep Point
As described in
"Mapping Samples to sweep Points with the Trace Detector"
on page 390, the number of samples taken during a sweep can be much larger than
the number of sweep points that are displayed in the measurement trace.
To determine the x-value of the sweep point, two different methods are available:
●
Start/stop
●
Bin-centered
Start/stop
This is the default (legacy) method for trace values in the frequency domain. The x-
value of the first sweep point corresponds to the starting point of the full measurement
span. The x-value of the last sweep point corresponds to the end point of the full mea-
surement span. All other sweep points are divided evenly between the first and last
points. The distance between two sweep points is span/(<no_sweep_points> -1).
Bin-centered
This is the default method for all marker values. The full measurement span is divided
by the number of sweep points. The result is the span that is evaluated for an individ-
ual sweep point, also referred to as a
bin
. The x-value of the sweep point is then
defined as the x-value at the center of the bin (bin/2).
Trace Configuration