![Ronde & Schwarz 1304.0004K03 User Manual Download Page 975](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987975.webp)
Remote Commands
R&S
®
FPL1000
975
User Manual 1178.3370.02 ─ 11
Example:
SYSTem:HPC HP
Manual operation:
See
SYSTem:IFGain:MODE
<Mode>
Configures the internal IF gain settings in HP emulation mode due to the application
needs. This setting is only taken into account for resolution bandwidth < 300 kHz and is
only available if a HP language is selected using
Parameters:
<Mode>
NORMal | PULSe
NORMal
Optimized for high dynamic range, overload limit is close to ref-
erence level.
PULSe
Optimized for pulsed signals, overload limit up to 10 dB above
reference level.
*RST:
NORM
Example:
SYST:IFG:MODE PULS
Manual operation:
See
SYSTem:LANGuage
<Language>
This command selects the system language.
Example:
SYST:LANG 'PSA'
Emulates the PSA.
Manual operation:
See
SYSTem:PSA:WIDeband
<State>
This command defines which option is returned when the
*OPT?
query is executed,
depending on the state of the wideband option.
It is only available for PSA89600 emulation.
Parameters:
<State>
ON | OFF | HIGH
OFF
The option is indicated as "B7J"
ON
The 40
MHz wideband is used.
The option is indicated as "B7J, 140".
HIGH
The 80
MHz wideband is used.
The option is indicated as "B7J, 122".
*RST:
OFF
Emulating Other Instruments' Commands