![Ronde & Schwarz 1304.0004K03 User Manual Download Page 721](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987721.webp)
Remote Commands
R&S
®
FPL1000
721
User Manual 1178.3370.02 ─ 11
10.6.14.3
Configuring the EMI Final Test
The commands required to configure the EMI final test are described here.
Useful commands for configuring EMI final tests described elsewhere:
●
[SENSe:]BANDwidth[:RESolution]:TYPE
●
[SENSe:]BANDwidth[:RESolution]
Remote commands exclusive to configuring EMI final tests
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:PEAKsearch:AUTO
CALCulate<n>:DELTamarker<m>:FUNCtion:FMEasurement:PSEarch:AUTO
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:PSEarch:AUTO
..............................721
CALCulate<n>:DELTamarker<m>:FUNCtion:FMEasurement:PEAKsearch:AUTO
CALCulate<n>:DELTamarker<m>:FUNCtion:FMEasurement:DWELl
...................................721
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:DWELl
.......................................... 721
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:PEAKsearch:AUTO
<State>
CALCulate<n>:DELTamarker<m>:FUNCtion:FMEasurement:PSEarch:AUTO
<State>
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:PSEarch:AUTO
<State>
CALCulate<n>:DELTamarker<m>:FUNCtion:FMEasurement:PEAKsearch:AUTO
<State>
Suffix:
<n>
.
1..n
<m>
1..n
Parameters:
<State>
CALCulate<n>:DELTamarker<m>:FUNCtion:FMEasurement:DWELl
<Time>
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:DWELl
<Time>
This command defines the dwell time during the final measurement.
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Parameters:
<Time>
Range:
100 us to 100 s
*RST:
1 s
Default unit: s
Manual operation:
See
Configuring and Performing Measurements