User Manual 1178.3370.02 ─ 11
Parameters:
<Bandwidth>
refer to data sheet
*RST:
RBW: AUTO is set to ON; DBW: 3MHz
Default unit: Hz
Example:
BAND 1 MHz
Sets the resolution bandwidth to 1 MHz
Manual operation:
See
See
See
See
See
[SENSe:]BWIDth[:RESolution]:AUTO
<State>
[SENSe:]BANDwidth[:RESolution]:AUTO
<State>
This command couples and decouples the resolution bandwidth to the span.
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
BAND:AUTO OFF
Switches off the coupling of the resolution bandwidth to the
span.
Manual operation:
See
See
[SENSe:]BWIDth[:RESolution]:RATio
<Ratio>
[SENSe:]BANDwidth[:RESolution]:RATio
<Ratio>
This command defines the ratio between the resolution bandwidth (Hz) and the span
(Hz).
Note that the ratio defined with this remote command (RBW/span) is reciprocal to that
of the coupling ratio (span/RBW).
Parameters:
<Ratio>
Range:
0.0001 to 1
*RST:
0.01
Example:
BAND:RAT 0.1
Manual operation:
See
[SENSe:]BWIDth[:RESolution]:TYPE
<FilterType>
[SENSe:]BANDwidth[:RESolution]:TYPE
<FilterType>
This command selects the resolution filter type.
When you change the filter type, the command selects the next larger filter bandwidth if
the same bandwidth is unavailable for that filter.
Setting Basic Spectrum RF Measurement Parameters