![Ronde & Schwarz 1304.0004K03 User Manual Download Page 603](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987603.webp)
Remote Commands
R&S
®
FPL1000
603
User Manual 1178.3370.02 ─ 11
Usage:
Query only
*PCB
<Address>
Pass control back
Indicates the controller address to which remote control is returned after termination of
the triggered action.
Setting parameters:
<Address>
Range:
0 to 30
Usage:
Setting only
*PRE
<Value>
Parallel poll register enable
Sets parallel poll enable register to the indicated value. The query returns the contents
of the parallel poll enable register in decimal form.
Parameters:
<Value>
Range:
0 to 255
*PSC
<Action>
Power on status clear
Determines whether the contents of the
ENABle
registers are preserved or reset when
the instrument is switched on. Thus a service request can be triggered when the instru-
ment is switched on, if the status registers ESE and SRE are suitably configured. The
query reads out the contents of the "power-on-status-clear" flag.
Parameters:
<Action>
0 | 1
0
The contents of the status registers are preserved.
1
Resets the status registers.
*RST
Reset
Sets the instrument to a defined default status. The default settings are indicated in the
description of commands.
The command is equivalent to
SYSTem:PRESet
.
Usage:
Setting only
Common Commands