![Ronde & Schwarz 1304.0004K03 User Manual Download Page 765](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987765.webp)
Remote Commands
R&S
®
FPL1000
765
User Manual 1178.3370.02 ─ 11
Manual operation:
See
See
See
[SENSe:]SWEep:TIME:AUTO
<State>
This command couples and decouples the sweep time to the span and the resolution
and video bandwidths.
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
SWE:TIME:AUTO ON
Activates automatic sweep time.
Manual operation:
See
See
See
[SENSe:]SWEep:TYPE:USED
This command queries the sweep type if you have turned on automatic selection of the
sweep type.
Return values:
<Type>
SWE
Normal sweep
FFT
FFT mode
10.8.3
Configuring the Vertical Axis (Amplitude, Scaling)
The following commands are required to configure the amplitude and vertical axis set-
tings in a remote environment.
................................................................................................ 765
...................................................................................767
.....................................................................................768
10.8.3.1
Amplitude Settings
The tasks for manual configuration are described in
Chapter 6.5.2, "Amplitude Set-
Useful commands for amplitude configuration described elsewhere:
●
Setting Basic Spectrum RF Measurement Parameters