![Ronde & Schwarz 1304.0004K03 User Manual Download Page 792](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987792.webp)
Remote Commands
R&S
®
FPL1000
792
User Manual 1178.3370.02 ─ 11
Tracking Generator Commands
The following commands are required to configure an internal tracking generator.
Useful commands for tracking generators described elsewhere:
●
●
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RPOSition
●
●
SOURce<si>:POWer[:LEVel][:IMMediate][:AMPLitude]
●
SOURce<si>:POWer[:LEVel][:IMMediate]:OFFSet
Remote commands exclusive to tracking generators:
SOURce<si>:FREQuency:COUPling[:STATe]
....................................................................792
SOURce<si>:INTernal:FREQuency:OFFSet
......................................................................792
.........................................................................................793
[SENSe:]CORRection:COLLect[:ACQuire]
........................................................................793
....................................................................................... 794
..........................................................................................794
[SENSe:]CORRection:TRANsducer:GENerate
.................................................................. 795
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:RVALue
............................795
SOURce<si>:FREQuency:COUPling[:STATe]
<State>
Enables or disables the usage of the internal generator as a tracking generator. The
tracking generator is only available for frequency sweeps in the Spectrum application.
Suffix:
<si>
.
1..n
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the tracking generator off
ON | 1
Switches the tracking generator on
*RST:
0
Example:
SOUR:FREQ:COUP:STAT ON
Manual operation:
See
SOURce<si>:INTernal:FREQuency:OFFSet
<FreqOffset>
Defines an offset of the tracking generator signal to the analyzer frequency.
Suffix:
<si>
.
1..n
Parameters:
<FreqOffset>
Default unit: HZ
Setting Basic Spectrum RF Measurement Parameters