FSE CALCulate Subsystem
1065.6016.12
6.35
E-16
:CALCulate<1|2>:LIMit<1 to 8>:ACPower:ALTernate<1|2>:STATe ON | OFF
This command activates the limit check for the first / second alternate adjacent channel for adjacent
channel power measurements. Before, the limit check must be activated using
CALC:LIM:ACP ON
.
Examples:
":CALC:LIM:ACP:ALT2:STAT ON"
Features:
*RST value:
OFF
SCPI:
device-specific
Modes:
A, VA
The numeric suffixes <1|2> or <1 to 8> are not significant for this command.
:CALCulate<1|2>:LIMit<1 to 8>:ACPower:ALTernate<1|2>:RESult?
This command queries the result of the limit check for the first / second alternate adjacent channel for
adjacent channel power measurements.
Parameter:
The result is returned in the form <result>, <result> where
<result> = PASSED | FAILED and where the first (second) returned value
denotes the lower (upper) alternate adjacent channel.
Examples:
":CALC:LIM:ACP:ALT2:RES?"
Features:
*RST value:
--
SCPI:
device-specific
Modes:
A, VA
This command is a query and therefore not assigned a *RST value. If the power measurement of the
adjacent channel is switched off, the command triggers a query error.
The numeric suffixes <1|2> or <1 to 8> are not significant for this command.
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......