![Rohde & Schwarz 1065.6000.20 Operating Manual Download Page 81](http://html1.mh-extra.com/html/rohde-and-schwarz/1065-6000-20/1065-6000-20_operating-manual_1477758081.webp)
CALCulate Subsystem FSE
1065.6016.12
6.24
E-16
:CALCulate<1|2>:LIMit<1 to 8>:CONTrol:SPACing LINear | LOGarithmic
This command makes a selection between linear and logarithmic interpolation for determining the
limit line from the frequency points.
Example:
":CALC:LIM:CONT:SPAC LIN"
Features:
*RST value:
LIN
SCPI:
device-specific
Modes:
A, VA, BTS, MS
:CALCulate<1|2>:LIMit<1 to 8>:UPPer[:DATA] <numeric_value>,<numeric_value>..
This command defines the values for the upper limit lines.
Example:
":CALC:LIM:UPP -10,0,0,-10
"
Features:
*RST value:
- (LIMit:STATe is set to OFF)
SCPI:
conforming
Modes:
A, VA, BTS, MS
The number of values for the CONTrol axis and the corresponding UPPer limit line have to be
identical. The unit must be identical with the unit selected by command
CALC:LIM:UNIT
.
If the measured values exceed the UPPer limit line, the limit check signals errors.
The unit must be identical with the unit selected by
CALC:LIM:UNIT
.
The units DEG, RAD, S, and HZ are available in the vector signal analysis mode only.
:CALCulate<1|2>:LIMit<1 to 8>:UPPer:STATe ON | OFF
This command defines the selected limit line as upper limit line.
Example:
":CALC:LIM:UPPer:STAT ON
"
Features:
*RST value:
OFF
SCPI:
conforming
Modes:
A, VA, BTS, MS
The limit check is switched on with command
CALCulate:LIMit:STATe ON
. The result of the limit
check can be queried with
CALCulate:LIMit<1 to 8>:FAIL?
.
:CALCulate<1|2>:LIMit<1 to 8>:UPPer:OFFSet <numeric_value>
This command defines an offset for the Y-axis of the selected relative upper limit line.
Example:
":CALC:LIM:UPP:OFFS 3dB"
Features:
*RST value:
0
SCPI:
device-specific
Modes:
A, VA
:CALCulate<1|2>:LIMit<1 to 8>:UPPer:MARGin <numeric_value>
This command defines the margin of the selected upper limit line.
Example:
":CALC:LIM:UPP:MARG 10dB"
Features:
*RST value:
0
SCPI:
device-specific
Modes:
A, VA, BTS, MS
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......