![Rohde & Schwarz 1065.6000.20 Operating Manual Download Page 90](http://html1.mh-extra.com/html/rohde-and-schwarz/1065-6000-20/1065-6000-20_operating-manual_1477758090.webp)
FSE CALCulate Subsystem
1065.6016.12
6.33
E-16
:CALCulate<1|2>:LIMit<1 to 8>:SPURious:FAILs? TXBand | OTXBand | RXBand | IDLeband
This command queries the number of limit violations of the spurious emissions measurement.
Examples:
":CALC:LIM:SPUR:FAIL? OTXB"
Features:
*RST value:
--
SCPI:
device-specific
Modes:
BTS, MS
TXBand
TX-band
OTXBand
Not TX-band
RXBand
RX-band (option FSE-K11 only)
IDLeband
IDLeband (option FSE-K10 only)
This command is a query and therefore not assigned a *RST value. The numeric suffixes <1|2> or
<1 to 8> are not significant for this command.
:CALCulate<1|2>:LIMit<1 to 8>:MARGin 0 to 100DB
This command sets / changes the value of the margin (safe difference to the actual limit) for the limit
check.
Examples:
":CALC:LIM:MARG 6DB"
Features:
*RST value:
3DB
SCPI:
device-specific
Modes:
BTS, MS
The numeric suffixes <1|2> or <1 to 8> are not significant for this command.
:CALCulate<1|2>:LIMit<1 to 8>:ACPower[:STATe] ON | OFF
This command switches on and off the limit check for adjacent channel power measurements. The
commands
CALC:LIM:ACP:ACH:STAT
or
CALC:LIM:ACP:ALT:STAT
must be used in addition to
specify whether the limit check is to be performed for the upper/lower adjacent channel or for the
alternate adjacent channels.
Examples:
":CALC:LIM:ACP ON"
Features:
*RST value:
OFF
SCPI:
device-specific
Modes:
A, VA
The numeric suffixes <1|2> or <1 to 8> are not significant for this command.
:CALCulate<1|2>:LIMit<1 to 8>:ACPower:ACHannel 0 to 100 dB, 0 to 100 dB
This command defines the limit for the upper/lower adjacent channel for adjacent channel power
measurements.
Parameter:
The first (second) numeric value is the limit for the upper (lower) adjacent
channel.
Examples:
":CALC:LIM:ACP:ACH 30DB, 30DB"
Features:
*RST value:
0 dB
SCPI:
device-specific
Modes:
A, VA
The numeric suffixes <1|2> or <1 to 8> are not significant for this command.
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......