![Rohde & Schwarz 1065.6000.20 Operating Manual Download Page 108](http://html1.mh-extra.com/html/rohde-and-schwarz/1065-6000-20/1065-6000-20_operating-manual_1477758108.webp)
FSE CALCulate Subsystem
1065.6016.12
6.51
E-16
:CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:POWer:RESult?ACPower | CPOWer | OBANdwidth |
OBWidth | CN | CN0
This command queries the results of the power measurement (see also
CALCulate:MARKer:
FUNCtion:POWer:SELect
.)
Example:
":CALC:MARK:FUNC:POW:RES? OBW"
Features:
*RST value:
-
SCPI:
device-specific
Modes:
A, VA, BTS, MS
ACPower
adjacent channel power measurement; Results are output separated by commas
in the following order:
Power of main channel
Power of lower adjacent channel 1
Power of upper adjacent channel 1
Power of lower adjacent channel 2
Power of upper adjacent channel 2
...
The number of results depends on the number of adjacent channels selected.
With logarithmic scaling (RANGE LOG), the power is output in dBm, with linear
scaling (RANGE LIN dB or LIN %) in W. If SENSe:POWer:ACHannel:MODE REL
is selected, adjacent channel power is output in dB.
CPOWer
channel power measurement
With logarithmic scaling (RANGE LOG), the channel power is output in dBm, with
linear scaling (RANGE LIN dB or LIN %) in W.
OBANdwidth | OBWidth occupied bandwidth power measurement
The return value is the occupied bandwidth in Hz
CN
signal / noise power measurement
The return value is always in dB..
CN0
signal-/ noise power measurement based on 1Hz bandwidth
The return value is always in dB/Hz
This command is only a query which is why it is not assigned an *RST value.
:CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:POWer[:STATe] OFF
This command switches the power measurement off.
Example:
":CALC:MARK:FUNC:POW OFF"
Features:
*RST value:
-
SCPI:
device-specific
Modes:
A-F, VA-D
This command is an event which is why it is not assigned an *RST value.
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......