![Rohde & Schwarz 1065.6000.20 Operating Manual Download Page 403](http://html1.mh-extra.com/html/rohde-and-schwarz/1065-6000-20/1065-6000-20_operating-manual_1477758403.webp)
Contents - Maintenance and Instrument Interfaces
FSE
1065.6016.12
8.10
E-15
RSIB Interface
The RSIB interface is only available if the FSE is equipped with a controller option, FSE-B15. It allows
the control of FSE by Visual C++ and VisualBasic programs but also by using the Windows applications
WinWord and Excel. The functions for the programming of control applications are provided by DLLs
RSIB32.DLL
(for 32-bit applications) and
RSIB.DLL
(for 16-bit applications). Instruments with MS DOS
controller option are restricted to 16-bit applications and local control in a Windows enviroment.
A Unix operating system can be installed on an external computer in addition to a Windows operating
system. In this case, the control applications are created either in C or C++. The supported Unix
operating systems include:
•
Sun Solaris 2.6 Sparc Station
•
Sun Solaris 2.6 Intel Platform
•
Red Hat Linux 6.2 x86 Processors
Control applications may run locally on the measuring instrument as well as on a remote controller in a
network. With local control, the name
’@local’
is specified for link setup by means of function
RSDLLibfind()
. If
’@local’
is not specified, the
RSIB.DLL
interprets the name as an IP address
and attempts to set up a link to the device via the Winsock interface.
Remote controller
(Windows NT or Windows 95)
FSE
0
1
2
3
4
5
6
7
8
9
.
-
CLR
BA CK
GHz
MHz
kHz
Hz
EXP
-d Bm
V
s
dBm
mV
ms
dB
µV
µs
dB..
nV
ns
AF OUTPUT
PROBE POWER
PROBE / CODE
30 d Bm
+
DC 0 V
MAX
M ADE IN GER MAN Y
HOLD
ST EP
50
W
SYSTEM
CONFIGURATION
PRESET
CAL
DISPLAY
INFO
MODE
SETUP
SETT ING
START
HARDCOPY
STATUS
SRQ
REMOTE
LOCAL
FREQUENCY
LEVEL
DATA ENTRY
CENTER /
SPAN /
UNIT
ST ART
ST OP
RANGE
MARKER
LINES
NORMAL SEARCH
D LINES
DELTA
LIMIT S
MKR
TRACE
SWEEP
DATA VARIATION
TRIGGER
SWEEP /
1
2
3
4
RBW
VBW
SWT
MENU
MEMORY
CONFIG
SAVE
RECALL
INPUT
FREQ
ZOOM
SCAN
COUPLING /
RUN
GEN OUTPUT 50
W
RF INPUT 1
MAX
REF /
REM OTE
EMI TEST RECEIVER 20 Hz . . . 7 GHz ESI
1088.7490
.
.
07
RF INPUT 2 20Hz ... 1GHz
30 dBm
+
MAX
50
W
Windows Environment
To access the measuring instruments via the RSIB interface the DLL should be installed in the
corresponding directories:
•
RSIB.DLL
in Windows NT
system
directory or control application directory.
•
RSIB32.DLL
in Windows NT
system32
directory or control application directory.
The DLLs are already installed in the corresponding directories on the measuring instruments.
For the different programming languages, files exist containing the declarations for the DLL functions
and the error code definitions.
Instruments witht Windows NT controller:
Visual Basic (16 bit):
’RSIB.BAS’
(C:/R_S/Instr/RSIB)
Visual Basic (32 bit):
’RSIB32.BAS’
(C:/R_S/Instr/RSIB)
C:
’RSIB.H’
(C:/R_S/Instr/RSIB)
Remote
control via
network
16-bit app.
16-bit application
32-bit app.
32-bit application
Local
applications
on instrument
RSIB network interface
RSIB DLL
RSIB.DLL
RSIB32.DLL
RSIB32.DLL
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......