![Rohde & Schwarz 1065.6000.20 Operating Manual Download Page 415](http://html1.mh-extra.com/html/rohde-and-schwarz/1065-6000-20/1065-6000-20_operating-manual_1477758415.webp)
Contents - Maintenance and Instrument Interfaces
FSE
1065.6016.12
8.22
E-15
Printer Interface (LPT)
The 25-pin LPT connector on the rear panel of the FSE is provided for the connection of a printer.. The
LPT interface is compatible with the CENTRONICS printer interface.
1
14
13
25
D0
D1
D2
D3
D4
D5
D6
D7
ACK
BUSY
PE
SELECT
STROBE
AUTOFEED
ERROR
INIT
SELECT IN
GND
GND
GND
GND
GND
GND
GND
GND
Pin
Signal
Input (I)
Output (O)
Description
1
STROBE
O
Pulse for transmitting a data byte, min. 1µs pulse width
(active LOW)
2
D0
O
Data Line 0
3
D1
O
Data Line 1
4
D2
O
Data Line 2
5
D3
O
Data Line 3
6
D4
O
Data Line 4
7
D5
O
Data Line 5
8
D6
O
Data Line 6
9
D7
O
Data Line 7
10
ACK
I
Indicates that the printer is ready to receive the next byte.
(active LOW)
11
BUSY
I
Signal is active when the printer cannot accept data.
(active HIGH)
12
PE
I
Signal is active when the paper tray is empty.
(active HIGH)
13
SELECT
I
Signal is active when the printer is selected.
(active HIGH)
14
AUTOFEED
O
When signal is active, the printer automatically performs a
linefeed after each line.
(active LOW)
15
ERROR
I
This signal is high when the printer has no paper, is not
selected or has an error status.
(active LOW)
16
INIT
O
Initialise the printer.
(active LOW)
17
SELECT IN
O
If signal is active, the codes DC1/DC3 are ignored by the
printer.
(active LOW).
18 - 25
GND
Ground connection.
Fig. 8-6 Pin assignments for the LPT connector.
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......