FSE
Status Reporting System
1065.6016.12
5.29
E-16
STATus QUEStionable:POWer Register
This register comprises all information about possible overloads of the unit.
It can be queried with commands
STATus:QUEStionable :POWer:CONDition?
and "
STATus
:QUEStionable:POWer [:EVENt]?
.
Table 5-9
Meaning of bits in STATus:QUEStionable:POWer register
Bit No.
Meaning
0
OVERload (Screen A)
This bit is set if the RF input is overloaded. ’OVLD’ will then be displayed.
1
UNDerload (Screen A) - Option FSE-B7
This bit is set if, during measurements in vector analyzer mode without capture buffer used, the lower level limit
in the IF path is violated.
2
IF_OVerload (Screen A)
This bit is set if the IF path is overloaded. ’IFOVLD’ will then be displayed.
3
not used
4
not used
5
not used
6
not used
7
not used
8
OVERload (Screen B)
This bit is set if the RF input is overloaded. ’OVLD’ will then be displayed.
9
UNDerload (Screen B) - Option FSE-B7
This bit is set if, during measurements without capture buffer used, the lower level limit in the IF path is violated.
10
IF_OVerload (Screen B)
This bit is set if the IF path is overloaded. ’IFOVLD’ will then be displayed.
11
not used
12
not used
13
not used
14
not used
15
This bit is always 0.
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......