FSE READ-Subsystem
1065.6016.12
6.159
E-16
READ:SPECtrum Subsystem
This subsystem provides the commands for starting measurements in the GSM BTS (option FSE-K11)
and GSM MS (option FSE-K10) Analyzer mode, which are used to measure the power of the spectral
components due to modulation and switching (modulation spectrum, transient spectrum), and for
reading out the results subsequently.
COMMAND
PARAMETERS
UNIT
COMMENT
READ
:SPECtrum
:MODulation
[:ALL]?
:SWITching
[:ALL]?
--
--
Option FSE-K11 or FSE-K10
query only
query only
:READ:SPECtrum:MODulation[:ALL]?
This command starts the measurement of the modulation spectrum of the base station or mobile and
reads out the result. The measurement is performed in the currently set frequency range.
Parameter:
The result is read out as a list of partial ASCII result strings separated by ’,’ in the
following format:
<Index>,<Freq1>,<Freq2>,<Level>,<Limit>, <Abs/Rel>,<Status> [,
<Index>,<Freq1>,<Freq2>,<Level>,<Limit>, <Abs/Rel>,<Status>]...
where the part set in ’[...]’ characterizes a partial result string which can be repeated n times.
<Index>:
0,
if the partial result string characterizes a
measurement range.
current number <>0,
if the partial result string characterizes a
single limit excess.
<Freq1>:
Start frequency of the measurement range or frequency where
the limit is exceeded.
<Freq2>:
Stop frequency of the measurement range or frequency where
the measured range is exceeded. The value of <Freq2> is
equal to the value of <Freq1>, if either the measurement is
performed in the time domain or the partial result string contains
a limit excess.
<Level>:
Measured maximum level of the partial range or measured level
at the test point.
<Limit>:
Limit in the partial range or at the test point.
<Abs/Rel>:
ABS <Level> and <Limit> are in absolute units (dBm)
REL <Level> and <Limit> are in relative units (dB)
<Status>:
Result of the limit check in character data form:
PASSED
no limit exceeded
FAILED
limit exceeded
MARGIN
margin exceeded
EXC
limited excess characterized as an exception
The frequencies <Freq1> and <Freq2> are always absolute and not referred to the carrier frequency.
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......