SENSe Subsystem FSE
1065.6016.12
6.172
E-16
:[SENSe<1|2>:]CORRection[:STATe] ON | OFF
This command activates/deactivates normalization of the measurement results.
Example:
":CORR ON "
Features:
*RST value:
OFF
SCPI:
conforming
Mode:
A
This command is only valid in conjunction with option Tracking Generator.
:[SENSe<1|2>:]CORRection:METHod TRANsmission | REFLexion
This command selects the kind of measurement with active tracking generator
(transmission/reflexion).
Example:
":CORR:METH TRAN "
Features:
*RST value:
TRANsmission
SCPI:
device specific
Mode:
A
This command is only valid in conjunction with option Tracking Generator.
:[SENSe<1|2>:]CORRection:COLLect[:ACQuire] THRough | OPEN
This command selects the kind of measurement for the reference values of the normalization
(response calibration).
Example:
":CORR:COLL THR"
Features:
*RST value:
SCPI:
conforming
Mode:
A
THRough
"TRANsmission" mode: calibration with direct connection between tracking
generator and device input.
"REFLexion" mode:
calibration with short circuit at the input
OPEN
only valid in "REFLexion" mode:calibration with open input
This command is an event which is why it is not assigned an *RST value an a query. It is only valid in
conjunction with option Tracking Generator.
:[SENSe<1|2>:]CORRection:RECall
This command restores the instrument setting that was valid for the measurement of the reference
data.
Example:
":CORR:REC"
Features:
*RST value:
-
SCPI:
conforming
Mode:
A
This command is an event which is why it is not assigned an *RST value and a query. It is only valid
in conjunction with option Tracking Generator.
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......