![Rohde & Schwarz 1065.6000.20 Operating Manual Download Page 131](http://html1.mh-extra.com/html/rohde-and-schwarz/1065-6000-20/1065-6000-20_operating-manual_1477758131.webp)
CONFigure-Subsystem FSE
1065.6016.12
6.74
E-16
:CONFigure:BURSt:PFERror:CONDition NORMal | EXTReme
This command defines the conditions for phase-frequency measurement.
Example:
":CONF:BURS:PFER:COND EXTR"
Features:
*RST value:
NORMal
SCPI:
device-specific
Modes:
BTS, MS
:CONFigure:BURSt:MACCuracy[:IMMediate]
This command selects measurement of the measurement accuracy of the base station or mobile.
Example:
":CONF:BURS:MACC"
Features:
*RST value:
--
SCPI:
device-specific
Modes:
BTS, MS
This command is an event and thus has no query and no *RST value assigned.
It is available only in conjunction with option GSM BTS Analyzer FSE-K11 or option GSM MS
Analyzer, FSE-K10, and Vector Signal Analysis, FSE-B7.
:CONFigure:BURSt:MACCuracy:COUNt 1 to 1000
This command sets the number of bursts used for the determination of average and maximum value.
Example:
":CONF:BURS:MACC:COUN 100"
Features:
*RST value:
1
SCPI:
device-specific
Modes:
BTS, MS
It is available only in conjunction with option GSM BTS Analyzer FSE-K11 or option GSM MS
Analyzer, FSE-K10, and Vector Signal Analysis, FSE-B7.
:CONFigure:BURSt:MACCuracy:CONDition NORMal | EXTReme
This command defines the conditions for modulation accuracy measurement.
Example:
":CONF:BURS:MACC:COND EXTR"
Features:
*RST value:
NORMal
SCPI:
device-specific
Modes:
BTS, MS
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......