Switchover to Remote Control
FSE
1065.6016.12
5.4
E-16
Remote Control via RS-232-Interface
Setting the Transmission Parameters
To enable an error-free and correct data transmission, the parameters of the unit and the controller
should have the same setting. Parameters can be manually changed in menu SETUP-GENERAL
SETUP in table COM PORT 1/2 or via remote control using the command
SYSTem:COMMunicate:SERial1|2:...
.
The transmission parameters of the interfaces COM1 and COM2 are factory-set to the following values:
Instruments with Windows NT controller:
baudrate = 9600, data bits = 8, stop bits = 1, parity = NONE and owner = INSTRUMENT.
Manually:
Setting interface COM1|2
½
Call SETUP-GENERAL SETUP menu
½
Select desired baudrate, bits, stopbit, parity and protocoll in table
COM PORT 1/2.
½
Set owner to Instrument or INSTR and DOS in table COM PORT1/2 (with
option FSE-B15 only)
½
Terminate input using one of the unit keys (=ENTER).
Instruments with MS DOS controller or without controller:
baudrate = 9600, data bits = 8, stop bits = 1, parity = NONE, protocoll = NONE
and owner = INSTRUMENT.
Manually:
Setting interface COM1|2
½
Call SETUP-GENERAL SETUP menu
½
Select desired baudrate, bits, stopbit, parity and protocoll in table
COM PORT 1/2.
½
Set owner to Instrument or INSTR and DOS in table COM PORT1/2 (with
MS DOS option FSE-B15 only)
½
Terminate input using one of the unit keys (=ENTER).
Return to Manual Operation
Return to manual operation is possible via the front panel or via RS-232 interface.
Manually:
½
Press the LOCAL key.
Notes:
Before switchover, command processing must be completed as
otherwise switchover to remote control is effected immediately.
– The LOCAL key can be disabled by the universal command LLO
(see chapter 8) in order to prevent unintentional switchover. In this
case, switchover to manual mode is only possible via remote
control.
– The LOCAL key can be enabled again by sending the control
codes "@LOC" via RS-232 (see chapter 8).
Via RS-232:
...
V24puts(port, ’
@
LOC’);
Set instrument to manual operation.
...
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......