![Rohde & Schwarz 1065.6000.20 Operating Manual Download Page 138](http://html1.mh-extra.com/html/rohde-and-schwarz/1065-6000-20/1065-6000-20_operating-manual_1477758138.webp)
FSE CONFigure-Subsystem
1065.6016.12
6.81
E-16
:CONFigure[:MS]:POWer:COUPled ON | OFF
This command switches between user-defined (OFF) and standard-defined (ON) level values.
Example:
":CONF:POW:COUP ON"
Feature: *RST value:
ON
ON
standard
OFF
user-defined
SCPI:
device-specific
Mode:
MS
:CONFigure[:MS]:POWer:LEVel 0 to 31
This command defines the power control level of the mobile.
Example:
":CONF:POW:LEV 5"
Features:
*RST value:
2 (P-GSM Phase I/II, E-GSM, R-GSM, GSM850)
0 (DCS1800, PCS1900)
SCPI:
device-specific
Mode:
MS
:CONFigure[:MS]:POWer:LIMit <numeric_value>
This command defines the level for the selection of level-dependent limit lines.
Example:
":CONF:POW:LIM 65DBM"
Feature:
*RST value:
depending on standard
SCPI:
device-specific
Mode:
MS
This command is only available for the setting
:CONFigure[:MS]:POWer:COUPled OFF
.
:CONFigure[:MS]:POWer:EXPected <numeric_value>
This command enters directly the rated output level of the mobile.
Example:
":CONF:POW:EXP 43DBM"
Features:
*RST value:
GMSK:
39 dBm (P-GSM Phase I/II, E-GSM, R-GSM, GSM850)
30 dBm (DCS1800, PCS1900)
EDGE
33 dBm (P-GSM Phase I/II, E-GSM, R-GSM, GSM850)
30 dBm (DCS1800, PCS1900)
SCPI:
device-specific
Mode:
MS
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......