![Rohde & Schwarz 1065.6000.20 Operating Manual Download Page 163](http://html1.mh-extra.com/html/rohde-and-schwarz/1065-6000-20/1065-6000-20_operating-manual_1477758163.webp)
FETCh-Subsystem FSE
1065.6016.12
6.106
E-16
:FETCh:BURSt:POWer:ALL?
Carrier Power measurement (
:CONFigure:MS:POWer:SINGle:STATe OFF
):
This command reads out the results of all individual steps during the measurement of the output
power of the base station or mobile.
Parameter:
The result is output as an ASCII string in the following format:
<Static Power Ctrl>,<Dyn Power Ctrl>,<Rat-Level>,<Act-Level>, <Delta>,<Status> with
<Static Power Ctrl>:
current static power control level
<Dyn Power Ctrl>:
current dynamic power control level
<Rat-Level>:
Rated value for the current power control level acc. to
standard dBm
<Act-Level>:
measured power in dBm
<Delta>:
Difference between the measured power and the power at the previous
static/dynamic power control level.
<Status>:
Result of limit check in character data form:
PASSED
no limits exceeded
FAILED
limit exceeded
Example:
":FETC:BURS:POW:ALL?"
Result:
0,0,43,44.1,0,PASSED,1,0,41,42.5,1.6,PASSED,1,1,35,32.5,5.6,FAILED
Features:
*RST value:
--
SCPI:
device-specific
Modes:
BTS, MS
Carrier Power Individual measurement (
:CONFigure:MS:POWer:SINGle:STATe ON
)
This command reads out the results of all individual steps during the measurement of the output
power of the base station or mobile.
Parameter:
The result is output as an ASCII string in the following format:
<Static Power Ctrl>,<Dyn Power Ctrl>,Rat-Level>,<Act-Level>,<RBW>,<ARFCN>,<CF>,
<Attenuation>,<Number of bursts>,<Status>
<Static Power Ctrl>: current static power control level
<Dyn Power Ctrl>:
current dynamic power control level
<Rat-Level>:
Rated value for the current power control level acc. to
standard dBm
<Act-Level>:
measured power in dBm
<RBW>:
resolution Bandwidth in kHz
<ARFCN>:
channel number
<CF>:
carrier frequency in in Hz
<Att>:
external attenuation in dBm
<Number of burst>: number of bursts
<Status>:
result of limit check in character data form:
PASSED
no limits exceeded
FAILED
limit exceeded
Example:
"
:FETC:BURS:POW:ALL?"
Result:
0,3,37,20.6915,1000,2,8.904E+008,20,1,FAILED,
0,3,37,20.3597,1000,2,8.904E+008,20,1,FAILED
Features:
*RST value:
--
SCPI:
device-specific
Modes:
BTS, MS
If no measurement has been performed yet, a query error results. This command is a query only and
therefore has no *RST value assigned. It is available only if measurement of the phase/frequency
error is selected (see
:CONFigure:BURSt:PFERror
).
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......