:DISPlay[:WINDow<1|2>]:TRACe<1 to 4>:MODE:CWRite ON | OFF
This command selects continuous display of the measured values (continuous write).
Example:
":DISP:TRAC3:MODE:CWR ON"
Features:
*RST value:
OFF
SCPI:
device-specific
Mode:
VA
:DISPlay[:WINDow<1|2>]:TRACe<1 to 4>:MODE:ANALog ON | OFF
This command selects continuous display of the measured values in the analyzer mode (analog
trace).
Example:
":DISP:TRAC3:MODE:ANAL ON"
Features:
*RST value:
OFF
SCPI:
device-specific
Mode:
A
:DISPlay[:WINDow<1|2>]:TRACe<1 to 4>:MODE:HCONtinuous ON | OFF
This command defines whether the traces in the min hold/max hold mode are reset after some
definite parameter changes.
Example:
":DISP:TRAC3:MODE:HCON ON"
Features:
*RST value:
OFF
SCPI:
device-specific
Mode:
A
OFF
The traces are reset after some definite parameter changes.
ON
This mechanism is switched off.
In general, parameter changes require a restart of the measurement before results are evaluated
(e.g. with markers). For those changes that are known to require a new measurement (e.g.
modification of the span), the trace is automatically reset so that erroneous evaluations of previous
results are avoided. This mechanism can be switched off for those exceptional cases where the
described behavior is unwelcome.
The numeric suffix in WINDow<1|2> is not significant.
:DISPlay[:WINDow<1|2>]:TRACe<1 to 4>[:STATe] ON | OFF
This command switches the display of the corresponding trace on or off.
Example:
":DISP:TRAC3 ON"
Features:
*RST value:
ON for TRACe1, OFF for TRACe2 to 4
SCPI:
conforming
Modes:
A, VA, BTS, MS
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......