![Rohde & Schwarz 1065.6000.20 Operating Manual Download Page 143](http://html1.mh-extra.com/html/rohde-and-schwarz/1065-6000-20/1065-6000-20_operating-manual_1477758143.webp)
CONFigure-Subsystem FSE
1065.6016.12
6.86
E-16
:CONFigure:SPECtrum:MODulation:RANGe ARFCn | TXBand | RXBand | COMBined | DCSRx1800
| G8Rxband | PCSRx1900
This command selects the frequency range for the measurement.
Example:
":CONF:SPEC:MOD:RANG TXB"
Features:
*RST value:
ARFCn
SCPI:
device-specific
Modes:
BTS, MS
ARFCn
ARFCN
±
1.8 MHz
TXBand
TX-Band
RXBand
RX-Band
COMBined
ARFCN
±
1.8 MHz / TX-Band
DCSRx1800
RX band DCS 1800 (option FSE-K10 only)
G8Rxband
RX band GSM 850 (option FSE-K10 only)
PCSRx1900
RX band PCS 1900 (option FSE-K10 only)
:CONFigure:SPECtrum:MODulation:TGATe ON | OFF
This command switches on or off the operating mode gating for the TX band. For TGATe OFF, 8
active slots are presumed.
Example:
":CONF:SPEC:MOD:TGAT ON"
Features:
*RST value:
OFF
SCPI:
device-specific
Mode:
BTS
:CONFigure:SPECtrum:SWITching[:IMMediate]
This command selects measurement of the spectrum due to switching transients.
Example:
":CONF:SPEC:SWIT"
Features:
*RST value:
--
SCPI:
device-specific
Modes:
BTS, MS
This command is an event and has thus no query and no *RST value assigned.
:CONFigure:SPECtrum:SWITching:COUNt 1 to 1000
This command defines the number of bursts used for determining the average and maximum values.
Example:
":CONF:SPEC:SWIT:COUN 100"
Features:
*RST value:
500 (GSM/DCS1800 phase I)
200 otherwise
SCPI:
device-specific
Modes:
BTS, MS
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......