![Rohde & Schwarz 1065.6000.20 Operating Manual Download Page 188](http://html1.mh-extra.com/html/rohde-and-schwarz/1065-6000-20/1065-6000-20_operating-manual_1477758188.webp)
FSE INSTrument Subsystem
1065.6016.12
6.131
E-16
:INSTrument<1|2>:COUPle NONE | MODE | X | Y | CONTrol | XY | XCONtrol | YCONtrol | ALL
This command defines the coupling between the two measurement windows screen A and B.
Example:
":INST:COUP NONE"
Features:
*RST value:
ALL
SCPI:
device specific
Modes:
A, VA
NONE
no coupling.
MODE
the operating mode of the two screens is coupled.
X or Y
the scaling of the x- or y axis of the two screens is coupled.
CONTrol
the trigger and gate parameter ,and the sweep parameters SINGle/ CONTinous
and COUNt of the two screens are coupled.
XY
the scaling of the x- and y-axis of the two screens are coupled.
XCONTrol bzw.
the trigger and gate parameter, and the sweep parameters SINGle CONTinous/
YCONTrol
COUNt of the two screens are coupled.
ALL
the scaling of the x- or y axis the trigger and gate parameter and the sweep
parameters SINGle/ CONTinous/ COUNt of the two screens are coupled.
The numeric suffix in INSTrument<1|2> is not significant.