![Rohde & Schwarz 1065.6000.20 Operating Manual Download Page 422](http://html1.mh-extra.com/html/rohde-and-schwarz/1065-6000-20/1065-6000-20_operating-manual_1477758422.webp)
FSE
List of Error Messages
1065.6016.12
9.1
E-1
9 List of Error Messages
The following list contains the error messages for errors occurring in the instrument. The meaning of
negative error codes is defined in SCPI, positive error codes mark errors specific of the instrument.
Error messages are entered in the error/event queue of the status reporting system in the remote
control mode and can be queried with the command
SYSTem:ERRor?
. The answer format of FSE to
the command is as follows:
<error code>, "<error text with queue query>; <remote control command concerned>"
The indication of the remote control command with prefixed semicolon is optional.
Example:
The command "
TEST:COMMAND
" generates the following answer to the query
SYSTem:ERRor? :
-113,"Undefined header;TEST:COMMAND"
The table contains the error code in the left-hand column. In the right-hand column the error text being
entered into the error/event queue or being displayed is printed in bold face. Below the error text, there
is an explanation as to the respective error.
SCPI-Specific Error Messages
No Error
Error code
Error text in the case of queue poll
Error explanation
0
No error
This message is output if the error queue does not contain any entries.
Command Error - Faulty command; sets bit 5 in the ESR register.
Error code
Error text in the case of queue poll
Error explanation
-100
Command Error
The command is faulty or invalid.
-101
Invalid Character
The command contains an invalid sign.
Example: A header contains an ampersand, "
SENSe&
".
-102
Syntax error
The command is invalid.
Example: The command contains block data the instrument does not accept.
-103
Invalid separator
The command contains an impermissible sign instead of a separator.
Example: A semicolon is missing after the command.
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......