SENSe Subsystem FSE
1065.6016.12
6.194
E-16
:[SENSe<1|2>:]FREQuency:CENTer:STEP:LINK SPAN | RBW | OFF
This command couples the step width of the center frequency to span (span >0) or to the resolution
bandwidth (span = 0) or cancels the couplings.
Example:
":FREQ:CENT:STEP:LINK SPAN"
Features:
*RST value:
SPAN
SCPI:
device-specific
Modes:
A, BTS, MS
:[SENSe<1|2>:]FREQuency:CENTer:STEP:LINK:FACTor 1 to 100 PCT
This command couples the step width of the center frequency with a factor to the span (span >0) or
to the resolution bandwidth (span = 0).
Example:
":FREQ:CENT:STEP:LINK:FACT 20PCT"
Features:
*RST value:
- (AUTO 0.1
×
SPAN is switched on)
SCPI:
device-specific
Modes:
A, BTS, MS
:[SENSe<1|2>:]FREQuency:SPAN 0 GHz to f
max
This command defines the frequency span of the analyzer.
Example:
":FREQ:SPAN 10MHz"
Features:
*RST value:
f
max
SCPI:
conforming
Mode:
A
The automatic coupling of the parameters is set to CENTER FIXED.
:[SENSe<1|2>:]FREQuency:SPAN:FULL
This command sets the maximum frequency span of the analyzer.
Example:
":FREQ:SPAN:FULL"
Features:
*RST value:
-
SCPI:
conforming
Mode:
A
This command is an event which is why it is not assigned an *RST value and has no query.
:[SENSe<1|2>:]FREQuency:SPAN:LINK CENTer | START | STOP
This command defines the coupling for frequency-span changes.
Example:
":FREQ:SPAN:LINK STOP"
Features:
*RST value:
CENTer
SCPI:
conforming
Mode:
A
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......