![Rohde & Schwarz 1065.6000.20 Operating Manual Download Page 73](http://html1.mh-extra.com/html/rohde-and-schwarz/1065-6000-20/1065-6000-20_operating-manual_1477758073.webp)
CALCulate Subsystem FSE
1065.6016.12
6.16
E-16
:CALCulate<1|2>:THReshold MINimum to MAXimum (depending on current unit)
This command defines the position of the thresholds.
Example:
":CALC:THR -82dBm"
Features:
*RST value:
- (STATe to OFF)
SCPI:
device-specific
Modes:
A, VA, BTS, MS
For marker scan functions MAX PEAK, NEXT PEAK etc., the threshold serves as the lowest limit for
maximum or minimum search. The units DEG, RAD, S, and HZ are only valid in conjunction with
option Vector Signal Analysis, FSE-B7.
:CALCulate<1|2>:THReshold:STATe ON | OFF
This command switches the threshold on or off.
Example:
":CALC:THR:STAT ON"
Features:
*RST value:
OFF
SCPI:
device-specific
Modes:
A, VA, BTS, MS
:CALCulate<1|2>:CTHReshold MINimum to MAXimum (depending on the current unit)
This command defines the position of a threshold line (base line), below which all measured values
are cleared.
Example:
":CALC:CTHR -82dBm"
Features:
*RST value:
- (STATe to OFF)
SCPI:
device-specific
Mode:
A, VA, BTS, MS
The units DEG, RAD, S, and HZ are only valid in conjunction with option Vector Signal Analysis,
FSE-B7.
:CALCulate<1|2>:CTHReshold:STATe ON | OFF
This command is for switching on or off the threshold line (base line), below which all measured
values are cleared.
Example:
":CALC:CTHR:STAT ON"
Features:
*RST value:
OFF
SCPI:
device-specific
Mode:
A, VA, BTS, MS
:CALCulate<1|2>:RLINe MINimum to MAXimum (depending on the current unit)
This command defines the position of the reference line.
Example:
":CALC:RLIN -10dBm"
Features:
*RST value:
- (STATe to OFF)
SCPI:
device-specific
Modes:
A, VA, BTS, MS
The reference line serves as a reference for the arithmetic operation of traces. The units DEG, RAD,
S, and HZ are only valid in conjunction with option Vector Signal Analysis, FSE-B7.
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......