FSE CALCulate Subsystem
1065.6016.12
6.45
E-16
:CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:ZOOM <numeric_value>
This command defines the range to be enlarged around the active marker.
Example:
":CALC:MARK:FUNC:ZOOM 1kHz"
Features:
*RST value:
-
SCPI:
device-specific
Mode:
A-F
The subsequent frequency sweep is stopped at the marker position and the frequency of the signal is
counted. This frequency becomes the new center frequency, the zoomed span is then set. This
command is an event which is why it is not assigned an *RST value and has no query.
:CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:NOISe[:STATe] ON | OFF
This command switches the noise measurement on or off.
Example:
":CALC:MARK:FUNC:NOIS ON"
Features:
*RST value:
OFF
SCPI:
device-specific
Mode:
A
The noise power density is measured at the position of the markers. The result can be queried with
CALCulate:MARKer:FUNCtion:NOISe:RESult?
.
:CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:NOISe:RESult?
This command queries the result of the noise measurement.
Example:
":CALC:MARK:FUNC:NOIS:RES?"
Features:
*RST value:
-
SCPI:
device-specific
Modes:
A, BTS, MS
This command is an event which is why it is not assigned an *RST value and has no query.
:CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:DEModulation:SELect AM | FM
This command selects the demodulation type.
Example:
":CALC:MARK:FUNC:DEM:SEL FM"
Features:
*RST value:
AM
SCPI:
device-specific
Mode:
A
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......