![Rohde & Schwarz 1065.6000.20 Operating Manual Download Page 259](http://html1.mh-extra.com/html/rohde-and-schwarz/1065-6000-20/1065-6000-20_operating-manual_1477758259.webp)
SENSe Subsystem FSE
1065.6016.12
6.202
E-16
:[SENSe<1|2>:]MSUMmary:REFerence <numeric_value>
This command selects the reference modulation.
Parameter:
<numeric_value> :=
0.001PCT to 1000PCT
for AM
0.1HZ to 10MHZ
for FM
0.0001RAD to 1000RAD for PM
Example:
":MSUM:REF 50PCT"
Features:
*RST value:
100PCTfor AM
100KHZfor FM
10RAD for PM
SCPI:
device-specific
Mode:
VA-A
:[SENSe<1|2>:]MSUMmary:REFerence:AUTO ONCE
This command sets the current absolute measured values of the main modulation signal as
reference values for the relative indication.
Example:
":MSUM:REF:AUTO ONCE"
Features:
*RST value:
--
SCPI:
device-specific
Mode:
VA-A
This command is an event and has therefore no query and no *RST value assigned.
:[SENSe<1|2>:]MSUMmary:MTIMe 0.1 s | 1 s
This command selects the measurement time for the summary markers.
Example:
":MSUM:MTIM 100US"
Features:
*RST value:
0.1S
SCPI:
device-specific
Mode:
VA-A