FSE CALCulate Subsystem
1065.6016.12
6.13
E-16
:CALCulate<1|2>:DELTamarker<1 to 4>:FUNCtion:FIXed[:STATe] ON | OFF
This command switches the relative measurement to a fixed reference value on or off.
Example:
":CALC:DELT:FUNC:FIX ON"
Features:
*RST value:
OFF
SCPI:
device-specific.
Modes:
A, VA-D, BTS, MS
The reference value is independent of the current trace.
:CALCulate<1|2>:DELTamarker<1 to 4>:FUNCtion:FIXed:RPOint:Y <numeric_value>
This command defines a new fixed reference value for the relative measurement.
Example:
":CALC:DELT:FUNC:FIX:RPO:Y -10dBm"
Features:
*RST value:
- (FUNction:FIXed[:STATe] is set to OFF)
SCPI:
device-specific
Modes:
A, VA
The reference value is independent of the current trace.
:CALCulate<1|2>:DELTamarker<1 to 4>:FUNCtion:FIXed:RPOint:Y:OFFSet <numeric_value>
This command defines an additional level offset for the relative measurement.
Example:
":CALC:DELT:FUNC:FIX:RPO:Y:OFFS 10dB"
Features:
*RST value:
0 dB
SCPI:
device-specific
Modes:
A, VA
The level offset is included in the output of the level value.
:CALCulate<1|2>:DELTamarker<1 to 4>:FUNCtion:FIXed:RPOint:X <numeric_value>
This command defines the new fixed reference frequency, time or symbols for the relative
measurement.
Example:
":CALC:DELT:FUNC:FIX:RPO:X 10.7MHz"
Features:
*RST value:
- (FUNction:FIXed[:STATe] is set to OFF)
SCPI:
device-specific
Mode:
A
The reference value is independent of the current trace. With span = 0, the reference time, otherwise
the reference frequency is defined.
:CALCulate<1|2>:DELTamarker<1 to 4>:FUNCtion:PNOise[:STATe] ON | OFF
This command switches the measurement of the phase noise on or off.
Example:
":CALC:DELT:FUNC:PNO ON"
Features:
*RST value:
OFF
SCPI:
device-specific
Mode:
A
When the phase noise is measured, the correction values for the bandwidth and the log amplifier are
automatically considered. The measurement uses the reference values defined by
FUNCtion:FIXed:RPOint:X
or
:Y
.
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......