:SYSTem:COMMunicate:GPIB[:SELF]:RTERminator LFEOI | EOI
This command changes the delimiter.
Example:
":SYST:COMM:GPIB:RTER EOI"
Features:
*RST value:
LFEOI
SCPI:
device-specific
Modes:
A, VA, BTS, MS
The instrument contains a DMA-channel for communication via IEC-bus. This ensures maximum
speed for the transfer of commands and data. The parser for command decoding integrated in the
instrument is, however, only activated by the delimiter when the command is completely transferred.
In order to make this possible for the transfer of binary data, too (e. g. trace data which are
retransferred into the instrument), the delimiter recognition must be switched to the EOI signal prior
to the transfer. Output of binary data from the instrument does not require such a switching.
:SYSTem:COMMunicate:GPIB:RDEVice<1|2>:ADDRess 0 to 30
This command changes the IEC/IEEE-bus address of the unit which is selected as hardcopy device
1 or 2, provided that the IEC/IEEE-bus interface of this unit is set as an interface.
Example:
":SYST:COMM:GPIB:RDEV2:ADDR 5"
Features:
*RST value:
4
SCPI:
conforming
Modes:
A, VA, BTS, MS
:SYSTem:COMMunicate:SERial<1|2>:CONTrol:DTR IBFull | OFF
:SYSTem:COMMunicate:SERial<1|2>:CONTrol:RTS IBFull | OFF
These commands switch the hardware handshake procedure for the given serial interface off (OFF)
or on (IBFull).
Examples:
":SYST:COMM:SER:CONT:DTR OFF"
":SYST:COMM:SER2:CONT:RTS IBF"
Features:
*RST value:
OFF
SCPI:
conforming
Modes:
A, VA, BTS, MS
The two commands have the same meaning. SERial1 and SERial 2 correspond to device COM1and
COM2, respectively.
:SYSTem:COMMunicate:SERial<1|2>[:RECeive]:BAUD 75 | 150 | 300 | 600 | 1200 | 2400 | 9600
This command sets the transmission speed for the given serial interface.
Example:
":SYST:COMM:SER:BAUD 2400"
Features:
*RST value:
9600
SCPI:
conforming
Modes:
A, VA, BTS, MS
SERial1 and SERial 2 correspond to device interface COM1 and COM2, respectively. Permissible
values are 75 Baud, 150 Baud, 300 Baud, 600 Baud, 1200 Baud, 2400 Baud, 4800 Baud, 9600
Baud.
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......