:[SENSe<1|2>:]DDEMod:SEARch:SYNC:PATTern <string>
This command defines the synchronization sequence.A file that has been set previously by the
command
DDEM:SEARch:SYNC:SELect
becomes invalid.
Example:
":DDEM:SEAR:SYNC:PATT "1101001"
Features:
*RST value:
""
SCPI:
device-specific
Mode:
VA-D
:[SENSe<1|2>:]DDEMod:SEARch:SYNC:STATe ON | OFF
This command switches the search for a synchronization sequence on or off.
Example:
":DDEM:SEARch:SYNC:STAT ON"
Features:
*RST value:
OFF
SCPI:
device-specific
Mode:
VA-D
:[SENSe<1|2>:]DDEMod:SEARch:SYNC:NAME <string>
This command selects a sync file to be edited or entered anew.
Example:
":DDEM:SEAR:SYNC:NAME "PATT_NEW"
Feature:
*RST value:
""
SCPI:
device-specific
Mode:
VA-D
:[SENSe<1|2>:]DDEMod:SEARch:SYNC:COMMent <string>
This command defines a comment on a sync file. This sync file must be previously selected by the
command
DDEM:SEARch:SYNC:NAME
.
Example:
":DDEM:SEAR:SYNC:COMM "PATTERN FOR PPSK"
Feature:
*RST value:
""
SCPI:
device-specific
Mode:
VA-D
:[SENSe<1|2>:]DDEMod:SEARch:SYNC:DATA <string>
This command defines a synchronization pattern for the sync file. Inputs other than "1" or "0" are
interpreted as "Don’t Care Bits". This sync file must be previously selected by the command
DDEM:SEARch:SYNC:NAME
.
Example:
":DDEM:SEAR:SYNC:DATA "1101001"
Feature:
*RST value:
""
SCPI:
device-specific
Mode:
VA-D
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......