![Rohde & Schwarz 1065.6000.20 Operating Manual Download Page 425](http://html1.mh-extra.com/html/rohde-and-schwarz/1065-6000-20/1065-6000-20_operating-manual_1477758425.webp)
List of Error Messages
FSE
1065.6016.12
9.4
E-1
Execution Error - Error on execution of a command; sets bit 4 in the ESR
register
Error code
Error text in the case of queue poll
Error explanation
-200
Execution error
Error on execution of the command.
-201
Invalid while in local
The command is not executable while the device is in local due to a hard local control.
Example: The device receives a command which would change the rotary knob state, but the device is in
local so the command can not be executed.
-202
Settings lost due to rtl
A setting associated with hard local control was lost when the device changed to LOCS from REMS or to
LWLS from RWLS.
-210
Trigger error
Error on triggering the device.
-211
Trigger ignored
The trigger (GET, *TRG or trigger signal) was ignored because of device timing considerations.
Example: The device was not ready to respond.
-212
Arm ignored
An arming signal was ignored by the device.
-213
Init ignored
Measurement initialisation was ignored as another measurement was already in progress.
-214
Trigger deadlock
The trigger source for the initiation of measurement is set to GET and subsequent measurement is
received. The measurement cannot be started until a GET is received, but the GET would cause an
interrupted-error)
-215
Arm deadlock
The trigger source for the initiation of measurement is set to GET and subsequent measurement is
received. The measurement cannot be started until a GET is received, but the GET would cause an
interrupted-error.
-220
Parameter error
The command contains a faulty or invalid parameter.
-221
Settings conflict
There is a conflict between setting of parameter value and instrument state.
-222
Data out of range
The parameter value lies out of the permissible range of the instrument.
-223
Too much data
The command contains too many data.
Example: The instrument does not have sufficient storage space.
-224
Illegal parameter value
The parameter value is invalid.
Example: The text parameter is invalid ,
TRIGger:SWEep:SOURce TASTe
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......