![Rohde & Schwarz 1065.6000.20 Operating Manual Download Page 141](http://html1.mh-extra.com/html/rohde-and-schwarz/1065-6000-20/1065-6000-20_operating-manual_1477758141.webp)
CONFigure-Subsystem FSE
1065.6016.12
6.84
E-16
:CONFigure[:MS]:SWEeptime STANdard | AUTO
This command selects the sweep-time computing mode for the spurious measurement:
Example:
":CONF:SWE AUTO"
Feature:
*RST value:
STANdard
SCPI:
device-specific
Mode:
MS
STANdard
The computation of the sweep time is based on a worst-case estimation
AUTO
The sweep time is reduced by a factor of 8 (assuming all slots are on).
:CONFigure[:MS]:MTYPe GMSK | EDGE
This command selects the standard (GMSK or EDGE).
Example:
":CONF:MTYP EDGE"
Features:
*RST value:
GMSK
SCPI:
device-specific
Mode:
MS