Continuation: Command Error
Error code
Error text in the case of queue poll
Error explanation
-138
Suffix not allowed
A suffix is not allowed for this command or at this position of the command.
Example: The command
*RCL
does not permit a suffix to be indicated.
-140
Character data error
The command contains a faulty text parameter
-141
Invalid character data
The text parameter either contains an invalid character or it is invalid for this command.
Example: Write error with parameter indication;
INPut:COUPling XC.
-144
Character data too long
The text parameter contains more than 12 characters.
-148
Character data not allowed
The text parameter is not allowed for this command or at this position of the command.
Example: The command
*RCL
requires a number to be indicated.
-150
String data error
The command contains a faulty string.
-151
Invalid string data
The command contains a faulty string.
Example: An END message has been received prior to the terminating apostrophe.
-158
String data not allowed
The command contains a valid string at a position which is not allowed.
Example: A text parameter is set in quotation marks,
INPut:COUPling
"DC"
-160
Block data error
The command contains faulty block data.
-161
Invalid block data
The command contains faulty block data.
Example: An END message was received prior to reception of the expected number of data.
-168
Block data not allowed
The command contains valid block data at an impermissible position.
Example: The command
*RCL
requires a number to be indicated.
-170
Expression error
The command contains an invalid mathematical expression.
-171
Invalid expression
The command contains an invalid mathematical expression.
Example: The expression contains mismatching parentheses.
-178
Expression data not allowed
The command contains a mathematical expression at an impermissible position.
-180
Macro error
A faulty macro has been defined, or an error has occurred during execution of a macro.
-181
Invalid outside macro definition
A macro parameter placeholder was encountered outside of a macro definition.
-183
Invalid inside macro definition
A macro definition is syntactically wrong
.
-184
Macro parameter error
A command inside the macro definition has the wrong number or type of parameters.
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......