![Rohde & Schwarz 1065.6000.20 Operating Manual Download Page 211](http://html1.mh-extra.com/html/rohde-and-schwarz/1065-6000-20/1065-6000-20_operating-manual_1477758211.webp)
READ-Subsystem FSE
1065.6016.12
6.154
E-16
:READ:BURSt:MACCuracy:RMS:STATus?
This command starts the measurement of the modulation accuracy of the base station or mobile and
reads out the status of the RMS-measurement taken over the selected number of bursts.
0: failed, 1: passed
Example:
":READ:BURS:MACC:RMS:STAT?"
Features:
*RST value:
--
SCPI:
device-specific
Modes:
BTS, MS
When the measurement is started the instrument automatically assumes the SINGLE mode.
An ongoing measurement can be aborted via the command
ABORt
. Further results of the modulation
accuracy measurement can be then queried without restart of the measurement via the
:FETCh:BURSt
-subsystem.
This command is a query only and therefore has no *RST value assigned. It is available only if
measurement of the modulation accuracy is selected (see
:CONFigure:BURSt:MACCuracy
).
:READ:BURSt:MACCuracy:RMS:AVERage?
This command starts the measurement of the modulation accuracy of the base station or mobile and
reads out the average of the RMS-measurement taken over the selected number of bursts.
Example:
":READ:BURS:MACC:RMS:AVER?"
Features:
*RST value:
--
SCPI:
device-specific
Modes:
BTS, MS
When the measurement is started the instrument automatically assumes the SINGLE mode.
An ongoing measurement can be aborted via the command
ABORt
. Further results of the modulation
accuracy measurement can be then queried without restart of the measurement via the
:FETCh:BURSt
-subsystem.
This command is a query only and therefore has no *RST value assigned. It is available only if
measurement of the modulation accuracy is selected (see
:CONFigure:BURSt:MACCuracy
).
:READ:BURSt:MACCuracy:RMS:MAXimum?
This command starts the measurement of the modulation accuracy of the base station or mobile and
reads out the maximum of the RMS-measurement for the selected number of bursts.
Example:
":READ:BURS:MACC:RMS:MAX?"
Features:
*RST value:
--
SCPI:
device-specific
Modes:
BTS, MS
When the measurement is started the instrument automatically assumes the SINGLE mode.
An ongoing measurement can be aborted via the command
ABORt
. Further results of the modulation
accuracy measurement can be then queried without restart of the measurement via the
:FETCh:BURSt
-subsystem.
This command is a query only and therefore has no *RST value assigned. It is available only if
measurement of the modulation accuracy is selected (see
:CONFigure:BURSt:MACCuracy
).
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......