![Rohde & Schwarz 1065.6000.20 Operating Manual Download Page 206](http://html1.mh-extra.com/html/rohde-and-schwarz/1065-6000-20/1065-6000-20_operating-manual_1477758206.webp)
FSE READ-Subsystem
1065.6016.12
6.149
E-16
:READ:BURSt:FERRor:MAXimum?
This command starts the measurement of the phase and frequency error of the base station or
mobile and reads out the maximum of the frequency error for the selected number of bursts.
Example:
":READ:BURS:FERR:MAX?"
Features:
*RST value:
--
SCPI:
device-specific
Modes:
BTS, MS
When the measurement is started the instrument automatically assumes the SINGLE mode.
An ongoing measurement can be aborted via the command
ABORt
. Further results of the
phase/frequency error measurement can be then queried without restart of the measurement via the
:FETCh:BURSt
-subsystem.
This command is a query only and therefore has no *RST value assigned. It is available only if
measurement of the phase/frequency error is selected (see
:CONFigure:BURSt:PFERror
).
:READ:BURSt:POWer?
Carrier Power measurement: (
:CONFigure:MS:POWer:SINGle:STATe OFF
)
This command starts the measurement of the maximum output power of the base station or mobile
and reads out the result.
Measurement of the maximum output power marks the beginning of a measurement cycle where
subsequently the limits of the static and dynamic power control levels are checked step by step
(
READ:BURSt:STATic?
or
READ:BURSt:DYNamic?
).
Parameter:
The result is read out as an ASCII string in the following format:
<Static Power Ctrl>,<Dyn Power Ctrl>,<Rat-Level>,<Act-Level>,
<Delta>,<Status>
<Static Power Ctrl>:
0
<Dyn Power Ctrl>:
0
<Rat-Level>:
rated value for the current power control level acc.
to standard in dBm
<Act-Level>:
measured power in dBm
<Delta>:
0
<Status>:
result of limit check in character data form:
PASSED
no limits exceeded
FAILED
limit exceeded
Example:
":READ:BURS:POW?"
Result:
0,0,43,44.1,0,PASSED
Features:
*RST value:
--
SCPI:
device-specific
Modes:
BTS, MS
When the measurement is started any ongoing measurement cycle is aborted.
An ongoing measurement can be aborted with the command
ABORT
. This command is a query only
and has therefore no *RST value assigned. It is available only when measurement of the maximum
carrier power is selected (see
:CONFigure:BURSt:POWer
).
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......