DISPlay Subsystem FSE
1065.6016.12
6.98
E-16
:DISPlay[:WINDow<1|2>]:TRACe<1 to 4>:Y[:SCALe]:RPOSition 0 to 100PCT
This command defines the position of the reference value.
Example:
":DISP:TRAC:Y:RPOS 50PCT"
Features:
*RST value:
100PCT (tracking generator)
50PCT (vector analyzer)
SCPI:
conforming
Modes:
A, VA
The numeric suffix in TRACe<1 to 4> is not significant. This command is only valid in conjunction
with option Tracking Generator or in vector analyzer mode.
:DISPlay[:WINDow<1|2>]:TRACe<1 to 4>:Y[:SCALe]:PDIVision <numeric_value>
This command defines the scaling of the Y-axis in the current unit.
Example:
":DISP:TRAC:Y:PDIV +1.20"
Features:
*RST value:
-
SCPI:
conforming
Mode:
VA
The numeric suffix in TRACe<1 to 4> is not significant.
:DISPlay[:WINDow<1|2>]:TRACe<1 to 4>:Y:SPACing LINear | LOGarithmic | PERCent
This command toggles between linear and logarithmic display.
Example:
":DISP:TRAC:Y:SPAC LIN"
Features:
*RST value:
LOGarithmic
SCPI:
conforming
Mode:
A
The linear display can be LIN/% (PERCent) or LIN/dB (LINear). The numeric suffix in TRACe<1 to 4>
is not significant.
:DISPlay[:WINDow<1|2>]:TRACe<1 to 4>:MODE WRITe | VIEW | AVERage | MAXHold | MINHold
This command defines the type of display and the evaluation of the traces.
Example:
":DISP:TRAC3:MODE MAXH"
Features:
*RST value:
WRITe for TRACe1, STATe OFF for TRACe2 to 4
SCPI:
device-specific
Modes:
A, VA, BTS, MS
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......