READ-Subsystem FSE
1065.6016.12
6.160
E-16
Example:
":READ:SPEC:MOD?"
Result:
0,890E6,915E6,-87.4,-108.0,ABS,FAILED,
1,893.2E6,893.2E6,-83.2,-108.0,ABS,FAILED,
2,895.7E6,895.7E6,-87.4,-108.0,ABS,FAILED
Features:
*RST value:
--
SCPI:
device-specific
Modes:
BTS, MS
The command
ABORt
aborts an ongoing measurement.
This command is only a query and therefore has no *RST value assigned. It is available only if
measurement of the modulation spectrum is selected. (see
:CONFigure:SPECtrum:MODulation
).
:READ:SPECtrum:SWITching[:ALL]?
This command starts the measurement of the transient spectrum of the base station or mobile and
reads out the result.
Parameter:
The result is read out as a list of partial ASCII result strings separated by ’,’ in the
format used for
READ:SPECtrum:MODulation[:ALL]?
.
Example:
":READ:SPEC:SWIT?"
Result:
0,833.4E6,833.4E6,37.4,-36.0,ABS,MARGIN,
1,834.0E6,834.0E6,-35.2,-36.0,ABS,FAILED,
2,834.6E6,834.6E6,-74.3,-75.0,REL,FAILED
0,835.0E6,835.0E6,-65,0,-60.0,REL,PASSED
Features:
*RST value:
--
SCPI:
device-specific
Modes:
BTS, MS
The command
ABORt
aborts an ongoing measurement.
This command is only a query and therefore has no *RST value assigned. It is available only if
measurement of the transient spectrum is selected. (see
:CONFigure:SPECtrum:SWITCHing
).
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......