FSE CALCulate Subsystem
1065.6016.12
6.27
E-16
:CALCulate<1|2>:LIMit<1 to 8>:LOWer:SPACing LINear | LOGarithmic
This command makes a selection between linear and logarithmic interpolation for the lower limit line.
Example:
":CALC:LIM:LOW:SPAC LIN"
Features:
*RST value:
LIN
SCPI:
device-specific
Modes:
A, VA, BTS, MS
:CALCulate<1|2>:LIMit<1 to 8>:FAIL?
This command queries the result of the limit check.
Example:
":CALC:LIM:FAIL?"
Features:
*RST value:
-
SCPI:
conforming
Modes:
A, VA, BTS, MS
The result of the limit check responds with 0 in case of PASS and with 1 in case of FAIL.
For measurements spectrum due to modulation and spectrum due to transients (options FSE-K10
and FSE-K11), the result of the limit check is queried with this command in frequency sweep mode.
:CALCulate<1|2>:LIMit<1 to 8>:CLEar[:IMMediate]
This command deletes the result of the current limit check.
Example:
":CALC:LIM:CLE"
Features:
*RST value:
-
SCPI:
conforming
Modes:
A, VA, BTS, MS
This command is an event which is why it is not assigned an *RST value.
:CALCulate<1|2>:LIMit<1 to 8>:COMMent <string>
This command defines a comment for the limit line selected.
Example:
":CALC:LIM:COMM ’Upper limit for spectrum’"
Features:
*RST value:
blank comment
SCPI:
device-specific
Modes:
A, VA, BTS, MS
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......