SENSe Subsystem FSE
1065.6016.12
6.196
E-16
:[SENSe<1|2>:]FREQuency:MODE CW | FIXed | SWEep
This command switches between frequency (SWEep) and time (CW | FIXed) domain in the analyzer
mode.
Example:
":FREQ:MODE SWE"
Features:
*RST value:
CW
SCPI:
conforming
Mode:
A
For CW and FIXed, the frequency setting is via command
FREQuency:CENTer
. In the SWEep
mode, the setting is via commands
FREQuency:STARt
,
STOP
,
CENTer
and
SPAN
.
:[SENSe<1|2>:]FREQuency:OFFSet <numeric_value>
This command defines the frequency offset of the instrument.
Example:
":FREQ:OFFS 1GHZ"
Features:
*RST value:
0 Hz
SCPI:
conforming
Modes:
A, VA