![Rohde & Schwarz 1065.6000.20 Operating Manual Download Page 101](http://html1.mh-extra.com/html/rohde-and-schwarz/1065-6000-20/1065-6000-20_operating-manual_1477758101.webp)
CALCulate Subsystem FSE
1065.6016.12
6.44
E-16
:CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:NDBDown <numeric_value>
This command defines the "N dB Down" value.
Example:
":CALC:MARK:FUNC:NDBD 3dB"
Features:
*RST value:
6dB
SCPI:
device-specific
Mode:
A
The temporary markers T1 and T2 are positioned by n dB below the active reference marker. The
frequency spacing of these markers can be queried with
CALCulate:MARKer:FUNCtion:
NDBDown:RESult
?.
:CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:NDBDown:STATe ON | OFF
This command switches the "N dB Down" function on or off.
Example:
":CALC:MARK:FUNC:NDBD:STAT ON"
Features:
*RST value:
OFF
SCPI:
device-specific
Modes:
A, BTS, MS
:CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:NDBDown:RESult?
This command queries the frequency spacing (bandwidth) of the "N dB Down" markers.
Example:
":CALC:MARK:FUNC:NDBD:RES?"
Features:
*RST value:
-
SCPI:
device-specific
Modes:
A, BTS, MS
This command is only a query which is why it is not assigned an *RST value.
:CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:NDBDown:FREQuency?
This command queries the frequencies of the "N dB Down" marker.
Example:
":CALC:MARK:FUNC:NDBD:FREQ?"
Features:
*RST value:
-
SCPI:
device-specific
Modes:
A, BTS, MS
The two frequency values are separated by comma and indicated in ascending order. This command
is only a query which is why it is not assigned an *RST value.
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......