List of Error Messages
FSE
1065.6016.12
9.2
E-1
Continuation: Command Error
Error code
Error text in the case of queue poll
Error explanation
-104
Data type error
The command contains an invalid value indication.
Example: ON is indicated instead of a numeric value for frequency setting.
-105
GET not allowed
A Group Execute Trigger (GET) is within a command line.
-108
Parameter not allowed
The command contains too many parameters.
Example: Command
SENSe:FREQuency:CENTer
permits only one frequency indication.
-109
Missing parameter
The command contains too few parameters.
Example: The command
SENSe:FREQuency:CENTer
requires a frequency indication.
-110
Command header error
The header of the command is faulty.
-111
Header separator error
The header contains an impermissible separator.
Example: the header is not followed by a "White Space", "
*ESE255
"
-112
Program mnemonic too long
The header contains more than 12 characters.
-113
Undefined header
The header is not defined for the instrument.
Example:
*XYZ
is undefined for every instrument.
-114
Header suffix out of range
The header contains an impermissible numeric suffix.
Example:
SENSe3
does not exist in the instrument.
-120
Numeric data error
The command contains a faulty numeric parameter.
-121
Invalid character in number
A number contains an invalid character.
Example: An "A" in a decimal number or a "9" in an octal number.
-123
Exponent too large
The absolute value of the exponent is greater than 32000.
-124
Too many digits
The number includes too many digits.
-128
Numeric data not allowed
The command includes a number which is not allowed at this position.
Example: The command
INPut:COUPling
requires indication of a text parameter.
-130
Suffix error
The command contains a faulty suffix.
-131
Invalid suffix
The suffix is invalid for this instrument.
Example: nHz is not defined.
-134
Suffix too long
The suffix contains more than 12 characters.
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......