FSE DIAGnostic-Subsystem
1065.6016.12
6.89
E-16
DIAGnostic Subsystem
The DIAGnostic subsystem contains the commands which support instrument diagnostics for
maintenance, service and repair. In accordance with the SCPI standard, all of these commands are
device-specific.
COMMAND
PARAMETERS
UNIT
COMMENT
DIAGnostic
:SERVice
:INPut
[:SELect]
:FUNCtion
:NSOurce
:INFO
:CCOunt
:ATTenuation<1|2|3>?
CALibration | RF
<numeric_value>,>numeric_value>
<Boolean>
--
--
--
no query
query only
:DIAGnostic:SERVice:INPut[:SELect] CALibration | RF
This command toggles between the RF input on the front panel and the internal 120-MHz reference
signal.
Example:
":DIAG:SERV:INP CAL"
Features:
*RST value:
RF
SCPI:
device-specific
Modes:
A, VA
:DIAGnostic:SERVice:FUNCtion <numeric_value>,<numeric_value>...
This command activates a service function.
Example:
":DIAG:SERV:FUNC 2,0,2,12,1"
Features:
*RST value:
-
SCPI:
device-specific
Modes:
A, VA
The service function is selected via five parameters: functional group number, board number,
function number, parameter 1 and parameter 2 (see service manual).
:DIAGnostic:SERVice:NSOurce ON | OFF
This command switches the 28-V supply at the rear connector of the noise source on and off.
Example:
":DIAG:SERV:NSO ON"
Features:
*RST value:
OFF
SCPI:
device-specific
Modes:
A, VA
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......