![Rohde & Schwarz 1065.6000.20 Operating Manual Download Page 200](http://html1.mh-extra.com/html/rohde-and-schwarz/1065-6000-20/1065-6000-20_operating-manual_1477758200.webp)
FSE OUTPut Subsystem
1065.6016.12
6.143
E-16
OUTPut Subsystem
The OUTPut subsystem checks the output features of the instrument.
In conjunction with option tracking generator, in the split screen mode, a distinction is made between
OUTPut1 (screen A) and OUTPut2 (screen B).
COMMAND
PARAMETERS
UNIT
COMMENT
OUTPut<1|2>
[:STATe]
:UPORt<1|2>
[:VALue]
:STATe
:AF
:SENSitivity
<boolean>
<binary>
<boolean>
<numeric_value
--
--
--
PCT |
HZ | KHZ |
DEG | RAD
Option Tracking Generator
Vector Signal Analysis
:OUTPut<1|2>[:STATe] ON | OFF
This command switches the tracking generator on or off.
Example:
":OUTP ON"
Features:
*RST value:
-
SCPI:
conforming
Modes:
A, VA
This command is only valid in conjunction with one of the options tracking generator.
:OUTPut<1|2>UPORt<1|2>[:VALue] #B00000000 to #B11111111
This command sets the control lines of the user ports.
Example:
":OUTP:UPOR2 #B10100101"
Features:
*RST value:
-
SCPI:
device-specific
Modes:
A, VA, BTS, MS
User port 1 or 2 is written with the given binary pattern. If the user port is programmed to INPut
instead of OUTPut, the output value is temporarily stored.
:OUTPut<1|2>UPORt<1|2>:STATe ON | OFF
This command switches the control line of the user ports between INPut and OUTPut.
Example:
":OUTP:UPOR:STAT ON"
Features:
*RST value:
OFF
SCPI:
device-specific
Modes:
A, VA, BTS, MS
With ON, the user port is switched to OUTPut, with OFF to INPut.
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......